Mouse IL-17A ELISA Kit (EM0008) is an ELISA kit with ultra sensitivity which can detect mouse Interleukin 17A up to 0.25 pg/ml, it is suitable for research use
- Ready to use ELISA kit based on sandwich immunoassay.
- Total time required is around 80min.
- Highly specific to mouse IL-17A.
- Applicable for serum and plasma samples.
Product | Mouse IL-17A ELISA Kit |
---|---|
Product # | EM0008 |
Category | ELISA Kits |
Species | Mouse |
Unit Size | 96T |
Storage | 2-8 degree Celsius. Room temperature with ice pack for shipment. |
Format | Readily available commercial kit |
Shelf-life | At least 12 months when stored properly. Opened kit shall be used in 30days. |
Application | For testing of mouse IL-17A in plasma, serum and/or other biological samples. |
Application Notes | This product is only for research use. |
Interleukin-17A (IL-17A) is a cytokine produced by T helper 17 (Th17) cells and plays important roles in the development of inflammatory diseases. Although IL-17F is highly homologous to IL-17A and binds the same receptor, the functional roles of this molecule remain largely unknown.
Mouse IL-17A ELISA Kit Principle
This Mouse IL-17A ELISA Kit is based on sandwich ELISA, the IL-17A specific monoclonal antibody was pre-coated in the microwells, the IL-17A in sample or the IL-17A standard will be captured by the coated antibody, after wash step, the unbonded IL-17A will be removed, and then another IL-17A specific monoclonal antibody is added and incubated. After then, the plate is washed again, and the substrate is added. The color of the wells is positively proportional to the IL-17A in it. After adding the stop solution, OD values are measured with ELISA reader. Unknown samples can be calculated from the standard curve.
Kit Components
This Mouse IL-17A ELISA kit is a readily commercial product, it has the following components, which is sufficient for 42 testing in duplicated wells. Please always check the kit instruction manual for packaging details.
- Microtiter plate, 96wells,
- Standard, will be diluted to 7 vials,
- HRP conjugate, 1 vial,
- TMB substrate, 1 vial,
- Assay buffer, 1 vial,
- Wash buffer, 1 vial,
- Sealer film or plate cover, 1p pc
- Stop solution, 1 vial,
Specificity and sensitivity
The sensitivity is 1.46pg/mL as MDD. The specificity is over 99%.
Result determination
The sample OD value is measured by an ELISA reader, then the standard curve is plotted. We recommend 4PL regression model or another equivalent model.
Storage and expiry
2-8 degree Celsius for at least 12 months.
Caution
- This kit is for research use only.
- The standard solution in the kit is mouse IL-17A .
- Please read the kit instruction manual carefully before testing.
- Each run will take 80min or more, please schedule your time before experiment.